Dexter Research Center, Inc.: We're everywhere it matters...

Jump to content

http://www.dexterresearch.com/home/products/thermopile-products/multi-channel-thermopiles/10-channel-themopile/


10 Channel

Thin Film Based Thermopile Detector

Features: A ten-channel thin-film thermopile in a TO-8 package. Each active area is 3.16mm x 0.4mm and offers low noise voltage. Internal aperture minimizes channel-to-channel crosstalk while increasing sensitivity.

Applications: The perfect single package choice for analyzing multiple anesthetic gases in demanding environments such as hospital operating rooms. Potential applications in non-contact temperature and spectral discrimination of materials.

Benefit: Ten-channel spectral differentiation for demanding applications.

Options:

  1. See Standard Windows and Filters (PDF) for list of optical filter options.
  2. Internal 30kΩ 5% NTC chip thermistor provides ambient package temperature measurement. See DC-4005 section of Thermistor Options (PDF).
  3. See Thermopile Configuration Table for more options.

Datasheet: For more detailed information see the datasheet (PDF).

10 Channel Thermopile detector circuit overlay

10 Channel Detector circuit overlay
(click image to see larger image)

10 Channel Thermopile Detector

10 Channel Thermopile Detector

10 Channel Thermopile Detector Technical Specifications

Specifications apply at 23°C with KBr Window and Nitrogen encapsulating gas

Parameter
Min
Typical
Max
Symbol
Units
Comments

Active Area size

3.16 x .4

AA

mm

Hot junction size, per element.

Element Area

1.264

A

mm2

 

Number of Junctions

40

 

 

Per element.

Number of Channels

10

 

 

Per detector package.

Output Voltage

90

115

130

Vs

μV

DC,  H=330μW/cm2 (3)

Signal-to-Noise Ratio

6,429

10,088

16,049

SNR

√Hz

DC, SNR=Vs/Vn

Responsivity

21.6

27.6

31.2

V/W

DC, =Vs/HA  (2)

Resistance

4.0

8.0

12

R

Detector element

Temperature Coefficient of

 

-.36

 

 

%/°C

Best linear fit, 0° to 85°C (1)

Temperature Coefficient of  R

 

-.2

 

 

%/°C

Best fit, 0° to 85°C (1)

Noise Voltage

8.1

11.4

14.0

Vn

nV/√Hz

Vn2=4kTR

Noise Equivalent Power

.26

.42

.65

NEP

nW/√Hz

DC,  NEP= VnHA/Vs (2)

Detectivity

1.7

2.7

4.3

D*

108√Hz/W

DC,  D*=Vs/ VnH√A  (2)

Time Constant

 

38

 

Τ

ms

Chopped, -3dB point (1)

Field of View

N/A

FOV

Degrees

See Assembly Drawings for FOV Description.

Package Type

TO-8 w/12 Pins

   

Standard package hole size:Ø.437

Element Matching

 

10

20

M

%

M=|VA-VB|/VB (2)

Element Separation

 

1.8 & 4.28

 

 

mm

Center to Center

Operating Temperature

-50

 

100

Ta

°C

 

General Specifications: 
Flat spectral response from 100nm to > 100µm.  Linear signal output from 10-6 to 0.1W/cm2
Maximum incident radiance 0.1W/cm2 , damage threshold ≥ .5W/cm2
Notes:      
(1)
Parameter is not 100% tested. 90% of all units meet these specifications.
(2)
A is detector area in cm2.
(3)