+1 (734) 426-3921
Questions about Dexter detectors? Call today. Outside the U.S., contact the representative nearest you.
3M

Thin Film Based Thermopile Detector
Specifications apply at 23°C with KBr Window and Argon encapsulating gas.
Parameter |
Min |
Typical |
Max |
Symbol |
Units |
Comments |
|
Active Area size |
3 x 3 |
AA |
mm |
Hot junction size, per element. |
||
|
Element Area |
9 |
A |
mm2 |
|||
|
Number of Junctions |
72 |
|
|
Per element. |
||
|
Number of Channels |
1 |
|
|
Per detector package. |
||
|
Output Voltage |
400 |
440 |
500 |
Vs |
μV |
DC, H=330μW/cm2 (3) |
|
Signal-to-Noise Ratio |
18,100 |
25,581 |
31,847 |
SNR |
√Hz |
DC, SNR=Vs/Vn |
|
Responsivity |
13.5 |
14.8 |
16.8 |
|
V/W |
DC, |
|
Resistance |
15 |
18 |
30 |
R |
kΩ |
Detector element |
|
Temperature Coefficient of |
|
-.36 |
|
|
%/°C |
Best linear fit, 0° to 85°C (1) |
|
Temperature Coefficient of R |
|
-.2 |
|
|
%/°C |
Best fit, 0° to 85°C (1) |
|
Noise Voltage |
15.7 |
17.2 |
22.1 |
Vn |
nV/√Hz |
Vn2=4kTR |
|
Noise Equivalent Power |
.93 |
1.16 |
1.64 |
NEP |
nW/√Hz |
DC, NEP= VnHA/Vs (2) |
|
Detectivity |
1.8 |
2.6 |
3.2 |
D* |
108√Hz/W |
DC, D*=Vs/ VnH√A (2) |
|
Time Constant |
|
100 |
|
Τ |
ms |
Chopped, -3dB point (1) |
|
Field of View |
63°/120° |
FOV |
Degrees |
See Assembly Drawings for FOV Description. |
||
|
Package Type |
TO-8 |
Standard package hole size:Ø.080 |
||||
|
Operating Temperature |
-50 |
|
100 |
Ta |
°C |
|
|
General Specifications: Notes: |
||||||