+1 (734) 426-3921
Questions about Dexter detectors? Call today. Outside the U.S., contact the representative nearest you.

6M

6M Detector circuit overlay
6M Detector circuit overlay
6M Thermopile Detector
6M Thermopile Detector

Thin Film Based Thermopile Detector

Features:
A thin film-based single element thermopile with the largest active area in our line up at 6.0mm diameter in a large TO-8 package
Applications:
Very large active area and high output makes it an excellent choice for laser power measurements.
Benefit:
Large active area and high output that is higher in cost and in a large package size.
Options:
  1. See Standard Windows and Filters (PDF) for list of optical filter options.
  2. See Thermopile Configuration Table for more options.
Datasheet:
For more detailed information see the datasheet
6M (PDF)

Specifications apply at 23°C with KBr Window and Argon encapsulating gas.

Parameter
Min
Typical
Max
Symbol
Units
Comments

Active Area size

Ø6mm Dia.

AA

mm

Hot junction size, per element.

Element Area

28.3

A

mm2

 

Number of Junctions

59

 

 

Per element.

Number of Channels

1

 

 

Per detector package.

Output Voltage

300

370

450

Vs

μV

DC,  H=330μW/cm2 (3)

Signal-to-Noise Ratio

12,552

18,317

28,662

SNR

√Hz

DC, SNR=Vs/Vn

Responsivity

3.2

4.0

4.8

V/W

DC, =Vs/HA  (2)

Resistance

15

25

35

R

Detector element

Temperature Coefficient of

 

-.36

 

 

%/°C

Best linear fit, 0° to 85°C (1)

Temperature Coefficient of  R

 

-.2

 

 

%/°C

Best fit, 0° to 85°C (1)

Noise Voltage

15.7

20.2

23.9

Vn

nV/√Hz

Vn2=4kTR

Noise Equivalent Power

3.25

5.10

7.45

NEP

nW/√Hz

DC,  NEP= VnHA/Vs (2)

Detectivity

.7

1.0

1.6

D*

108√Hz/W

DC,  D*=Vs/ VnH√A  (2)

Time Constant

 

221

 

Τ

ms

Chopped, -3dB point (1)

Field of View

75°/137°

FOV

Degrees

See Assembly Drawings for FOV Description.

Package Type

TO-8

   

Standard package hole size:Ø.437

Operating Temperature

-50

 

100

Ta

°C

 

General Specifications: 
Flat spectral response from 100nm to > 100µm.  Linear signal output from 10-6 to 0.1W/cm2
Maximum incident radiance 0.1W/cm2 , damage threshold ≥ .5W/cm2

Notes:      
(1) Parameter is not 100% tested. 90% of all units meet these specifications.
(2) A is detector area in cm2.
(3)Test Conditions: 500K Blackbody source; Detector active surface 10cm from 0.6513cm Diameter Blackbody Aperture.