+1 (734) 426-3921
Questions about Dexter detectors? Call today. Outside the U.S., contact the representative nearest you.

M14

M14 Detector circuit overlay
M14 Detector circuit overlay
M14 Thermopile Detector
M14 Thermopile Detector

Thin Film Based Thermopile Detector

Features:
A single-channel thin film-based thermopile offering very low noise in a compact TO-18 package with a small active area of 0.92mm x 0.4mm. This is one of the lowest noise thermopiles you can buy and provides a fast 14ms time constant with Argon encapsulation gas.
Applications:
Excellent for ear thermometers and handheld non-contact temperature measurement.
Benefit:
Small package size and very low noise with lower signal-to-noise ratio.
Options:
  1. See Standard Windows and Filters (PDF) for list of optical filter options.
  2. See Thermopile Configuration Table for more options.
Datasheet:
For more detailed information see the datasheet
M14 (PDF)

Specifications apply at 23°C with KBr Window and Argon encapsulating gas.

Parameter
Min
Typical
Max
Symbol
Units
Comments

Active Area size

.92 x .4

AA

mm

Hot junction size, per element.

Element Area

.368

A

mm2

 

Number of Junctions

12

 

 

Per element.

Number of Channels

1

 

 

Per detector package.

Output Voltage

15

20

25

Vs

μV

DC,  H=330μW/cm2 (3)

Signal-to-Noise Ratio

1,852

2,857

5,000

SNR

√Hz

DC, SNR=Vs/Vn

Responsivity

12.4

16.5

20.6

V/W

DC, =Vs/HA  (2)

Resistance

1.5

3.0

4.0

R

Detector element

Temperature Coefficient of

 

-.36

 

 

%/°C

Best linear fit, 0° to 85°C (1)

Temperature Coefficient of  R

 

-.2

 

 

%/°C

Best fit, 0° to 85°C (1)

Noise Voltage

5.0

7.0

8.1

Vn

nV/√Hz

Vn2=4kTR

Noise Equivalent Power

.24

.43

.66

NEP

nW/√Hz

DC,  NEP= VnHA/Vs (2)

Detectivity

.93

1.4

2.5

D*

108√Hz/W

DC,  D*=Vs/ VnH√A  (2)

Time Constant

 

14

 

Τ

ms

Chopped, -3dB point (1)

Field of View

53°/105°

FOV

Degrees

See Assembly Drawings for FOV Description.

Package Type

TO-18

   

Standard package hole size:Ø.080

Operating Temperature

-50

 

100

Ta

°C

 

General Specifications: 
Flat spectral response from 100nm to > 100µm.  Linear signal output from 10-6 to 0.1W/cm2
Maximum incident radiance 0.1W/cm2
, damage threshold ≥ .5W/cm2

Notes:      
(1) Parameter is not 100% tested. 90% of all units meet these specifications.
(2) A is detector area in cm2.
(3)Test Conditions: 500K Blackbody source; Detector active surface 10cm from 0.6513cm Diameter Blackbody Aperture.