+1 (734) 426-3921
Questions about Dexter detectors? Call today. Outside the U.S., contact the representative nearest you.

ST60 LCC

ST60 LCC Detector circuit overlay
ST60 LCC Detector circuit overlay
ST60 LCC Thermopile Detector
ST60 LCC Thermopile Detector

Silicon Based Thermopile Detector

Features:
A single-channel silicon-based thermopile in a LCC package for surface mount assembly in a small active area of 0.61mm x 0.61mm. Time constant of 18ms with Nitrogen encapsulation gas. Delivers a very low Temperature Coefficient of Responsivity of -0.04%/°C. This detector has a very short thermal shock response to ambient temperature chang
Applications:
Excellent for non-contact temperature, and horizon sensor.
Benefit:
Low cost surface mount assembly and small active area size with medium output.
Options:
  1. See Standard Windows and Filters (PDF) for list of optical filter options.
  2. ST60R LCC version offers a low-cost (20% tolerance) poly-silicon resistor to be used as a PTC thermistor.
  3. Internal 30kΩ 5% NTC chip thermistor provides ambient package temperature measurement. See DC-4005 section of Thermistor Options (PDF).
  4. Internal aperture precisely defines active area for applications with FOV and/or spot size requirements. See Aperture Options (PDF) for available sizes.
  5. See Thermopile Configuration Table for more options.
Datasheet:
For more detailed information see the datasheet
ST60_LCC (PDF)

Specifications apply at 23°C with KBr Window and Nitrogen encapsulating gas.

Parameter
Min
Typical
Max
Symbol
Units
Comments

Active Area size

.61 x .61

AA

mm

Hot junction size, per element.

Element Area

.37

A

mm2

 

Number of Junctions

80

 

 

Per element.

Number of Channels

1

 

 

Per detector package.

Output Voltage

50

62

74

Vs

μV

DC,  H=330μW/cm2 (3)

Signal-to-Noise Ratio

1,381

1,981

2,891

SNR

√Hz

DC, SNR=Vs/Vn

Responsivity

40.7

50.5

60.3

V/W

DC, =Vs/HA  (2)

Resistance

40

60

80

R

Detector element

Temperature Coefficient of

 

-.04

 

 

%/°C

Best linear fit, 0° to 85°C (1)

Temperature Coefficient of  R

 

.11

 

 

%/°C

Best fit, 0° to 85°C (1)

Noise Voltage

25.6

31.3

36.2

Vn

nV/√Hz

Vn2=4kTR

Noise Equivalent Power

.42

.62

.89

NEP

nW/√Hz

DC,  NEP= VnHA/Vs (2)

Detectivity

.69

.98

1.44

D*

108√Hz/W

DC,  D*=Vs/ VnH√A  (2)

Time Constant

 

18

 

Τ

ms

Chopped, -3dB point (1)

Field of View

45°/81°

FOV

Degrees

See Assembly Drawings for FOV Description.

Package Type

LCC

   

Standard package hole size: Ø.069

Operating Temperature

-50

 

100

Ta

°C

 

ST60R Thermistor Options

~24

30

~36

RT

PTC Poly-Silicon resistor on detector die.

ST60R Thermistor Temperature Coefficient of R

.107

.11

.113

 

%/°C

ΔR/(RΔT), Best fit, 0° to 85°C (1)

General Specifications: 
Flat spectral response from 100nm to > 100µm.  Linear signal output from 10-6 to 0.1W/cm2
Maximum incident radiance 0.1W/cm2 , damage threshold ≥ .5W/cm2
Notes:      
(1) Parameter is not 100% tested. 90% of all units meet these specifications.
(2)
A is detector area in cm2.
(3)
Test Conditions: 500K Blackbody source; Detector active surface 10cm from 0.6513cm Diameter Blackbody Aperture.